Entering Special Transducer Pipe Parameters A-7
December 2003
Entering Parameters for Clamp-On Transducers
(cont.)
8C. Entering the Outside Diameter of the Pipe
If you do not know the outside diameter (OD) of your pipe,
Sound Speeds and Pipe Size Data contains some standard pipe
sizes with their corresponding outside diameters, otherwise
measure it.
To enter the outside diameter:
1. Press the appropriate feature key to select the units.
2. Use the numeric keys to enter the outside diameter.
3. Press [ENT].
9C. Entering the Pipe Wall Thickness
If you are not sure of the pipe wall thickness, Sound Speeds and
Pipe Size Data contains some standard pipe sizes with their
corresponding wall thicknesses. Also, the pipe wall thickness can
be determined by using the optional internal ultrasonic thickness
gage as described in Chapter 9, Using the Ultrasonic Thickness
Gage.
Use the numeric keys to enter the pipe wall thickness. Notice the
units are the same as selected in the SYSTeM option. Press
[ENT].